Juxta-Lite compression wrap Medium Short 28cm Length with Separate Medium AFW Catalog 23134017
FDA Recall Z-1696-2010 · Posted 2010-05-27
Advertisement
Recalling FirmCircaid Medical Products Inc
Recall StatusTerminated
Product CodeMDR
Date Initiated2010-04-20
Date Terminated2011-01-06
Product Quantity372 total, all sizes
Reason for Recall
Some systems in these lots have Built-In Pressure System lines with the incorrect amount of spacing. Some systems in these lots had I inch space between each line with no tension applied. The system should have 1.25 inches of space. The correct spacing (1.25") is necessary to achieve the measured tension ranges. The incorrect spacing may result in the inability to achieve the correct tension level
Firm Action
CircAid Medical Products Important Device Product Recall letter dated Apr 20, 2010 was mailed to consignees describing problem, listing affected product and distribution dates. Customers were instructed to locate and return these Juxta-Lite Systems from stock and notify their customers A Retum Authorization number and contact name was provided.
this listing for product returns.
Root Cause
Process control
Distribution Pattern
Worldwide distribution: USA, Japan, Canada, Netherlands