Terminated

Bard Ventralight" ST Mesh with Echo PS" 8"x10" Reorder Number: 5955810 Indicated for use in the reconstruction of soft tissue deficiencies, such as the repair of hernias.

FDA Recall Z-1696-2014 · Posted 2014-06-02

Recall StatusTerminated
Product CodeGCJ
Date Initiated2014-04-24
Date Terminated2015-12-23

Reason for Recall

Pouch holding sterile inflation assembly and inflation adapter may be open or have a weak seal.

Firm Action

Davol, Inc., Subs. C. R. Bard, Inc.sent an Urgent Medical Device Recall letter dated April 28, 2014, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to examine their inventory and quarantine product subject to the recall, contact Davol Customer Service at 1-800-556-6275 or C.R. Bard's Medical Services and Support Department at 1-800-562-0027 for instructions on how to return the product to Davol, Consignees were also instructed to complete and return the recall effectiveness check form, and notify their customers if the product was further distributed. For questions regarding this recall call 1-800-556-6275.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to AUSTRIA, BELGIUM, CANADA, CYPRUS, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ITALY, LUXEMBOURG, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, and UNITED KINGDOM