Alinity i Anti-TPO Reagent Kit, List Number 09P3521

FDA Recall Z-1696-2023 · Posted 2023-06-08

Recall StatusOpen, Classified
Product CodeJZO
Date Initiated2023-04-24
Product Quantity110 kits

Reason for Recall

Due to a manufacturing fill volume error, Alinity i Anti-TPO Reagent contains a marginally lower fill volume than detailed in the Kit Contents section of the Instructions for Use (IFU).

Firm Action

On April 24, 2023, Abbott issued a Product Correction letter, via FedEx priority overnight express, for this issue. Customers were instructed to provide a copy of the Product Correction letter to the laboratory manager, supervisor or health professional responsible for the impacted product and to complete and return the Customer Reply form.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the states of CA, CO, FL, GA, IL, KY, LA, MD, MI, MN, NJ, NY, OH, RI, SC, SD, TN, TX, WA. No international distribution of affected lot.