Terminated

CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic Radiosurgery System

FDA Recall Z-1697-2019 · Posted None

Recalling FirmAccuray Incorporated
Recall StatusTerminated
Product CodeIYE
Date Initiated2018-04-10
Date Terminated2021-04-01
Product Quantity43 units

Reason for Recall

There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.

Firm Action

The firm issued a field safety notice by letter on 04/10/2019. The letter explained the issue and provided safety instructions. All impacted customers with affected systems will be contacted to arrange for a software upgrade.

Root Cause

Other

Distribution Pattern

US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand