Urethral Warming System-Galil Medical Warming Catheter. The product is contained within prostate kits. This product is used in cryosurgery to prevent freezing of tissue.

FDA Recall Z-1698-2008 · Posted 2008-09-03

Recalling FirmGalil Medical
Recall StatusTerminated
Product CodeGEH
Date Initiated2007-12-11
Date Terminated2008-09-02
Product Quantity541 catheters

Reason for Recall

The warming catheter was manufactured with the inlet and outlet tubes switched.

Firm Action

The recalling firm telephoned their customers in April 2008 to inform them of the problem and the need to return the product. There customers were also asked to contact the firm if they had actually used any of the kits with the serial numbers listed, even if they experienced no problem. As a follow up to the telephone call, a letter was issued to the customers. For assistance, please call 770-565-6166.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution --- USA including states of PA, SC, FL, NY, TN, and WA, and country of Canada