Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min) Contains AH165 Circuit, AH290 Chamber Component: No. a conduit for respiratory gas between the patient and a ventilator.

FDA Recall Z-1698-2025 · Posted 2025-05-09

Recalling FirmVyaire Medical
Recall StatusOpen, Classified
Product CodeBZE
Date Initiated2025-04-10
Product Quantity3053

Reason for Recall

Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.

Firm Action

On April 10, 2025, the firm notified customers via emailed letters titled URGENT: MEDICAL DEVICE RECALL. Customers are instructed to examine their inventory for the affected lots of product. Product in inventory should be quarantined. For in-use product - 1) For affected product in use that is utilizing the adaptor connection, immediately stop use of the product. 2) For affected product in use that is not utilizing the adaptor connection, ventilator circuit may continue to be utilized. Customers were provided with additional precautions to take to ensure proper function. Customers can choose between returning or destroying affected product. Firm will provide replacement product. If you have any questions regarding this field action, please call AirLife at 1-800-433-2797, or e-mail at [email protected].

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Germany, Poland, Belgium, Spain, Romania, UK, Italy, Netherlands, India, Slovenia, Qatar, and Saudi Arabia.