Zimmer Allen medullary cement plugs, 1 - 20 dia. flange/10 mm dia. core, 1 - 24 mm dia. flange/12 mm dia. core (polyethylene with barium sulfate) with inserter, sterile, Zimmer, Warsaw, Indiana; Catatog Number: 00801102001. Total joint arthroplasty to control, restrict, or impede the flow of cement. Larger plugs are useful in revision surgery where a wide, smooth Intramedullary canal must be plugged.

FDA Recall Z-1699-2009 · Posted 2009-07-23

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeJDI
Date Initiated2009-06-08
Date Terminated2009-12-15
Product Quantity133

Reason for Recall

The seal on the sterile barrier pouch may be inadequate, resulting in lack of assurance of sterility.

Firm Action

Zimmer Orthopedic Surgical Products issued an "Urgent: Medical Device Recall" notice dated June 8, 2009 informing users of the affected devices and actions to be taken including the return of product and acknowledgement of receipt of notice by returning the Recall Certification Form by fax to 1-330-364-0974.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution -- United States, Canada and Japan.