Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Magic3 Go Intermittent Urinary Catheters, male Catalog Numbers: a) REF 50810G, b) REF 50812G, c) REF 50814G, d) REF 50816G, e) REF 50818G, f) REF 53810G, g) REF 53812G, h) REF 53814G, i) REF 53816G, j) REF 53818G, k) REF 53820G

FDA Recall Z-1699-2023 · Posted 2023-06-08

Recalling FirmC.R. Bard Inc
Recall StatusOpen, Classified
Product CodeEZD
Date Initiated2023-04-20
Product Quantity7427880 devices

Reason for Recall

Some catheters had a manufacturing defect where the clear resealing label separating from the product foil pouch, resulting in a possible opening into the product packaging that could impact product sterility.

Firm Action

The firm issued an URGENT MEDICAL DEVICE RECALL to its consignee beginning on 04/20/2023 by Fed Ex and email. The notice explained the problem, risk, and requested the following actions be taken: Discontinue distribution of the product and dispose of all affected product, Distributors are directed to notify their customers. For further assistance please contact: 1-844-8BD- LIFE (1-844-823-5433) Say "Recall" when prompted; M-F 8am - 5pm CT

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑