Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Zimmer Coonrad/Morrey total elbow cement restrictor with nozzle, 1 - 16 mm, 1 - 25 mm, sterile, Zimmer, Dover, Ohio; Catatog Number: 32810503800. Total joint arthroplasy to control, restrict or impede the flow of cement. Larger plugs are useful in revision surgery where a wide, smooth Imtramedullary canal must be plugged.

FDA Recall Z-1700-2009 · Posted 2009-07-23

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeJDI
Date Initiated2009-06-08
Date Terminated2009-12-15
Product Quantity110

Reason for Recall

The seal on the sterile barrier pouch may be inadequate, resulting in lack of assurance of sterility.

Firm Action

Zimmer Orthopedic Surgical Products issued an "Urgent: Medical Device Recall" notice dated June 8, 2009 informing users of the affected devices and actions to be taken including the return of product and acknowledgement of receipt of notice by returning the Recall Certification Form by fax to 1-330-364-0974.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution -- United States, Canada and Japan.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑