Magic3 Go Intermittent Urinary Catheters, pediatric Catalog Numbers: a) REF 52810G
FDA Recall Z-1700-2023 · Posted 2023-06-08
Advertisement
Recalling FirmC.R. Bard Inc
Recall StatusOpen, Classified
Product CodeEZD
Date Initiated2023-04-20
Product Quantity5760 devices
Reason for Recall
Some catheters had a manufacturing defect where the clear resealing label separating from the product foil pouch, resulting in a possible opening into the product packaging that could impact product sterility.
Firm Action
The firm issued an URGENT MEDICAL DEVICE RECALL to its consignee beginning on 04/20/2023 by Fed Ex and email. The notice explained the problem, risk, and requested the following actions be taken: Discontinue distribution of the product and dispose of all affected product, Distributors are directed to notify their customers.
For further assistance please contact: 1-844-8BD- LIFE (1-844-823-5433) Say "Recall" when prompted; M-F 8am - 5pm CT
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement