Vortex MEDICAL AngioVac Circuit, REF/UPN VTX3100, STERILE, Rx Only, Single Use Only -- The AngioVac Circuit and AngioVac Circuit Pack are designed to be used with the AngioVac Cannula.
FDA Recall Z-1701-2016 · Posted 2016-05-20
Recalling FirmAngiodynamics Inc. (Navilyst Medical Inc.)
Recall StatusTerminated
Product CodeDWE
Date Initiated2016-03-21
Date Terminated2017-09-18
Product QuantityDomestic: 339 kits
Reason for Recall
Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in packaged AngioVac Circuit Packs.
Firm Action
On 3/21/2016, AngioDynamics sent recall notification packages (dated 3/21/2016) to the consignees via Federal Express. Consignees are instructed to follow the instructions provided on the Maquet Cardiopulmonary recall letter (dated 2/23/2016) and return the Reply Verification Tracking Form, provided in the recall notification.
Root Cause
Environmental control
Distribution Pattern
Nationwide Distribution.