Various hysterectomy, vaginal packs
FDA Recall Z-1702-2022 · Posted None
Advertisement
Recalling FirmAmerican Contract Systems, Inc.
Recall StatusOpen, Classified
Product CodeOHD
Date Initiated2022-06-16
Product Quantity29 units
Reason for Recall
Product was exposed to multiple sterilization cycles without validation for multiple exposures.
Firm Action
The recalling firm began notifying customers of the recall beginning June 13, 2022. Customer notification was initially made via email and an "Urgent Medical Device Recall" letter dated June 16, 2022.
Customers are asked to review inventory and segregate affected product, which should be identified with the Bag Serial Number. Affected product should be destroyed, with evidence of destruction provided back to the recalling firm.
A Field Action Response Form should be completed and returned to [email protected] as soon as possible, even if the customer has no inventory.
If you have any questions, please call 816-920-5846.
Root Cause
Process change control
Distribution Pattern
Distribution within the United States to health systems and distributors in AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, and TX
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement