Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin with plasma activator for use in the determination of the activated partial thromboplastin time (APTT) and other coagulation procedures requiring an activated partial thromboplastin reagent.
FDA Recall Z-1703-2015 · Posted 2015-06-02
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeGFO
Date Initiated2015-03-25
Date Terminated2017-01-24
Product Quantity27501
Reason for Recall
Recovery for assayed controls are out of assigned ranges. A drift >3 seconds in the normal APTT range and/or >15% in the pathological range has been observed. Patient samples with recoveries close to the medical decision points could exhibit a deviation up to 4 seconds for the normal range and up to 33% for the pathological range.
Firm Action
On March 24, 2015 an Urgent Medical Device Recall letter was mailed to all affected customers to notify them about the issue, associated risks, and actions to be taken.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide Distribution. US naionwide including Puerto Rico, Barbados and Mexico.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.