syngo Workflow SLR; The information system syngo Workflow SLR is a digital radiology information system (RIS) with integrated modules for patient administration, examination, reporting, statistics and system administration. The system electronically displays stores, retrieves, transfers, exchanges, and prints.
FDA Recall Z-1704-2015 · Posted 2015-06-02
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeLMD
Date Initiated2015-05-06
Date Terminated2016-08-17
Product Quantity131
Reason for Recall
A potential exists for order transactions from interfaced HIS (Hospital Information System) systems to be lost during a restart of interface processes when using the Order Batching feature, leading to diagnosis delay.
Firm Action
Siemens sent an Safety Advisory Notice, dated May 6, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Customers were given work around instructions to follow and informed that a software fix would be implemented to permanently resolve the issue.
For further questions please call (610) 219-6300.
Root Cause
Software design
Distribution Pattern
Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.