Stryker Xia 3 Anti-Torque Key Manual Surgical Instrument; Catalog Number 48237026. Product Usage: : Reusable instrument used to aid in the final tightening of the spinal construct. Intended Use: The Anti-Torque Key allows the Torque Wrench to align with the tightening axis and maximizing the torque needed to lock the implant assembly during final tightening of the spinal construct.

FDA Recall Z-1706-2012 · Posted 2012-06-05

Recalling FirmStryker Spine
Recall StatusTerminated
Product CodeNKB
Date Initiated2012-01-12
Date Terminated2013-03-21
Product Quantity95 units

Reason for Recall

Stryker received four reports regarding a handle which separates at the point where it is welded to the shaft on the Xia 3 Anti-Torque Key during surgery.

Firm Action

Stryker Spine sent an Urgent Product Recall letter dated January 12, 2012 with an attached Customer Response forms via FEDEX to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine their inventory to identify the affected product and return using the prepaid mailing label provided. The letter asked customers to complete the Customer Response Form and fax to 201.760.8370. For questions call 201.760.8298.

Root Cause

Mixed-up of materials/components

Distribution Pattern

US Nationwide Distribution