Terminated

NxStage PureFlow B Solution, REF RFP-401, M535RPF4010 NxStage PureFlow B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis.

FDA Recall Z-1707-2019 · Posted None

Recalling FirmNxStage Medical, Inc.
Recall StatusTerminated
Product CodeKPO
Date Initiated2019-04-11
Date Terminated2020-06-08
Product Quantity3,940 units

Reason for Recall

Certain lots were mislabeled with incorrect product number.

Firm Action

On April 11, 2019, NxStage Medical Inc. issued URGENT MEDICAL DEVICE RECALL notices to customers via courier service. Customers were advised to take the following actions: 1. Check all boxes in your Pureflow B solution inventory. If you find any boxes or individual bags of Pureflow B solution from the affected lots listed in the table below, separate them from your other inventory and do not use them. Refer to Appendix A for pictures to assist in locating part or SKU and lot number of your Pureflow B solution inventory. 2. Promptly complete and return the online response form at the link www.novasyte.com/nxstageapr2019 to acknowledge your receipt of this notice, even if you do not have the affected product on hand. 3. Upon completion of the online response form, your facility will be scheduled for a product return pick up date based on the availability you provide. NxStage is partnered with Novasyte to assist in this action. Novasyte specializes in providing outsourced commercial service teams and technologies for the medical device industry. For any assistance regarding this action please contact Novasyte using the information below: Phone: (760) 205-3020 Email: [email protected]

Root Cause

Labeling mix-ups

Distribution Pattern

US Nationwide Distribution - AZ, CA, DC, FL, GA, IL, IN, MI, NC, NH, NV, PA, TX & VA