Terminated

Toshiba America Medical Systems (TAMS) Aquilion 64 CT system, computed tomography x-ray system.

FDA Recall Z-1708-2008 · Posted 2008-09-29

Recall StatusTerminated
Product CodeJAK
Date Initiated2007-10-01
Date Terminated2012-07-16
Product Quantity1203 total

Reason for Recall

Excessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used.

Firm Action

Toshiba sent a letter to customers in July 2007, and sent field service representatives to update the Data Aquisition System in September/October 2007.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution, including USA and Japan.