BD Nexiva Dual Port 20GA 1.00IN -Closed IV Catheter System intended for use in the peripheral vascular system Catalog Number: 383536
FDA Recall Z-1710-2019 · Posted None
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeFOZ
Date Initiated2019-03-01
Date Terminated2020-05-21
Product Quantity6,720 units
Reason for Recall
Leak from the septum assembly or the canister may burst under high pressure which could result in blood exposure to a health care worker.
Firm Action
BD MDS (Medication Delivery Solutions) contacted via email and provided a hard copy letter via FedEx on March 1, 2019. The tetter describes the problem , action to take:
Immediately review your inventory for the specific Catalog and lot number listed above.Destroy all product subject to the recall following your institutions process for destruction. Share this recall notification with all users of the product within your facility to ensure that they are also aware of this recall. . Complete the attached Customer Response Form and return to the BD contact noted on the
form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification and process your product replacement
BD is requesting the distributor to identify their customers and either notify their
customers regarding the recall or provide BD a customer list. BD will mail additional communications to the customers that are identified on the lists provided by the distributor.
Questions: Customer/Technical Support
888-237-2762 OPT 3, OPT 2
Monday Friday
between 8:00am and 5:00pm (EST
Root Cause
Release of Material/Component prior to receiving test results
Distribution Pattern
US distribution to one distributor in VA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.