SMR glenosphere impactor/extractor; Used to implant the glenosphere in the SMR Reverse Shoulder replacement.
FDA Recall Z-1711-2016 · Posted None
Advertisement
Recalling FirmLimacorporate S.p.A
Recall StatusTerminated
Product CodeKWS
Date Initiated2016-04-08
Date Terminated2016-12-05
Product Quantity314 units
Reason for Recall
Complaints of intra-operative breakage of the glenosphere impactors/extractors.
Firm Action
The affected firms received written notification to all distributors. The letter instructed the consignee to identify and remove all affected product in stock and return parts to Lima USA.
Root Cause
Component design/selection
Distribution Pattern
Distributed US (nationwide) and in Australia.