SJM Peel Away Introducer, REF 405153, Lot 2694252, Length 14 cm, 7F (2.33 mm), Max Guidewire o.d .038" (.965 mm), contents 5, 2011-08, Manufacturer St Jude Medical 14901 DeVeau Place, Minnetonka, MN 55345-2126 USA, Sterile EO, Rx only. Intended for use for the introduction of catheters into a vessel.

FDA Recall Z-1714-2009 · Posted 2009-08-17

Recalling FirmSt. Jude Medical
Recall StatusTerminated
Product CodeDQX
Date Initiated2009-06-24
Date Terminated2012-03-29
Product Quantity200

Reason for Recall

St. Jude Medical has initiated a voluntary recall of two lots of Peel Away Introducers . Isolated lots could contain some items that have been mis-labelled with the incorrect French size of the outer box label. Some 9F products have been labelled as 7F and vice versa. Only the outer box label is affected by this error, the labels on the individual product pouches correctly identify the product. St Jude Medical has not received any reports of injures resulting from this mislabelling to date.

Firm Action

The Consignee was provided with a St. Jude Medical "Urgent Product Information" letter and a doctor Acknowledgement form, via e-mail on June 2009. The letter described the problem and the product involved and advised to isolate and discontinue use of the product involved. The Consignee will notify the end user and remove any remaining affected units and will notify St Jude Medical of the results.

Root Cause

Labeling mix-ups

Distribution Pattern

OUS only: Belgium.