Terminated

The Stryker NAV3i Platform is a mobile cart platform consisting of the sub-components Stryker Nav3 Graphite Camera, SPC-3 Computer and IO-Tablet (touch screen) as well as the monitor, articulated arms and cart housing. This Navigation system is used in combination with surgical software for cranial, spine, ENT, orthopedic and trauma procedures.

FDA Recall Z-1715-2014 · Posted 2014-06-04

Recall StatusTerminated
Product CodeHAW
Date Initiated2014-05-13
Date Terminated2014-09-15
Product Quantity20

Reason for Recall

The Stryker NAV3i Platform Power Boxes produced between December 9, 2013 and March 7, 2014 may have been assembled without electrical insulation of the soldering joints

Firm Action

On May 13, 2014, Stryker Instruments sent URGENT MEDICAL DEVICE RECALL notification to their consignees. Notification included the description of the affected product, reason for the recall, risk to health, and actions to be taken by the Customer/User. Once Stryker receives the response from the customer, a Stryker Representative will contact the customer to set up a time to perform the upgrade to the affected power box. Old Power Boxes will be returned to Stryker and placed in an industrial compactor/recycler for destruction. Questions regarding the recall are directed to Kelly Jo Whipple M-F 8am-5pm @ 269-389-2921 or [email protected]

Root Cause

Process change control

Distribution Pattern

Distributed in the states of AR, CA, IN, LA, MA, MD, MI, MT, NV, and TX.