Philips Healthcare Ingenuity Elite Computed Tomography X-Ray System

FDA Recall Z-1719-2016 · Posted None

Recalling FirmPhilips Healthcare
Recall StatusTerminated
Product CodeJAK
Date Initiated2015-09-06
Date Terminated2021-03-04
Product QuantityUS - 423

Reason for Recall

Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

Firm Action

Philips plans to initiate a field corrective action that will install a new software release to address the issue. Philips will modify the systems in the field to correct the defect at no cost to the customer. 1. You will contact customers and initiate a software update to correct the defect. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call (1-800) 722-9377, Option 5

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

US Distribution - AZ, PA, FL, DC, W VA, NC, SC, AR, OK, OH, NY, NV, TX, PA, IN, CA, MN, MA, DC, MS, MD, OR, IL, CO, MO, VT, IA, AL, LA, GA and WA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.