Ambu aScope" 5 Broncho HD 5.6/2.8 Sampler Set. Model Number: 622002000US
FDA Recall Z-1723-2025 · Posted 2025-05-06
Advertisement
Recalling FirmAmbu Inc.
Recall StatusOpen, Classified
Product CodeEOQ
Date Initiated2025-04-07
Product Quantity160 units
Reason for Recall
Incorrect labeling in which the front red pouch label did not match the actual size of the medical device. The label on the front of the scope incorrectly states 5.0/2.2 while the back label, shipping box and product inside correctly states 5.6/2.8.
Firm Action
Ambu notified sales representative first on 04/07/2025 with an "Urgent Field Safety Notice" letter and they are instructed to visit the affected accounts to notify them of the recall. The letter instructs customers to identify and quarantine any affected units on hand, to arrange for the return of affected units or destroy them on site, and to complete and return the response form.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of CA, CO, FL, GA, KY, MD, MT, NJ, NV, PA, TN, TX, VA.