Terminated

Thoracentesis Set, RPN C-THS-850, Order # G03286

FDA Recall Z-1725-2019 · Posted None

Recalling FirmCook Inc.
Recall StatusTerminated
Product CodePYC
Date Initiated2019-04-09
Date Terminated2020-04-28
Product Quantity5 units

Reason for Recall

Affected products may have been manufactured with the catheter distal end hole too small

Firm Action

On April 9, 2019 Cook Medical issued URGENT MEDICAL DEVICE RECALL notices to customers via courier service. Customers were advised to take the following Actions: 1. Examine your inventory immediately to determine if you have affected product(s), and quarantine affected product(s). Immediately cease all distribution and use of these products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. NOTE: Unaffected products that are returned will not be credited. 3. Complete and return the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return it via fax (812.339.7316) or by email to [email protected]. 4. The notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30 am and 5:00 pm (Eastern Time) or by email to [email protected].

Distribution Pattern

Domestically: OH; Internationally: CANADA, SPAIN, ITALY, BELGIUM, GERMANY, UNITED ARAB EMIRATES, GREECE, SWITZERLAND, OMAN, MONACO, SAUDI ARABIA, AUSTRALIA, HONG KONG, INDIA & KOREA

Official Source

View FDA Recall →