Alere Triage D-Dimer Test PN 98100, Lot W53884B. The Alere Triage D-Dimer Test is a fluorescence immunoassay to be used with the Alere Triage Meters for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA anticoagulated whole blood and plasma specimens. . The Alere Triage D-Dimer test is intended for use as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.

FDA Recall Z-1727-2014 · Posted 2014-06-06

Recalling FirmAlere San Diego, Inc.
Recall StatusTerminated
Product CodeGHH
Date Initiated2014-05-02
Date Terminated2015-02-27
Product Quantity373 kits

Reason for Recall

Alere initiated this recall because a limited number of Alere Triage D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sample these affected devices will either generate an error code (e.g. E4 error), or they may generate an incorrect result.

Firm Action

Alere sent an Urgent Medical Device Recall letter dated May 2, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are instructed to stop using all product from the affected lot, lot W53884B, and to discard any unused product. The customers were also instructed to complete and return the verification form indicating any credit or replacement product requested. They were also instructed to contact their primary consignees and advise them of the notice. Should you have any questions about the information contained in this notification, please call (877) 308-8287. and /or email:[email protected]

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide Distribution only.