Nikon Metrology XTH cabinet x-ray Systems which includes variations STH225, XTH225ST, and MCT225.

FDA Recall Z-1727-2020 · Posted None

Recalling FirmNikon Metrology
Recall StatusTerminated
Product CodeRCE
Date Initiated2019-12-23
Date Terminated2024-06-13
Product Quantity110 units

Reason for Recall

The manufacturer identified a risk that their products might fail to comply with the cabinet x-ray system performance standard limit on x-ray emissions [21 CFR 1020.40(c)(1)] due to an insufficient safety margin for the means of attachment of a piece of interior x-ray shielding. Double sided industrial tape was used for attachment - there have been only 2 instances of detachment and although the piece of shielding detached it remained mostly in place supported by wires that pass through that piece of shielding. The repair is accomplished by adding a bracket to fasten the piece of shielding securely to the inside of the x-ray system. Mfr reports that no systems were found to have exceeded the emission limit when checked prior to repair.

Firm Action

Nikon sent Notification letters to affected customers. The letter identified the Description of the Possible Defect, Potential Health Risk from the Defect, and actions to be taken. For question contact the service department via email at xray-support.nm-us@nikon or phone at (810)220-4360.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

US Nationwide Distribution