iGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.

FDA Recall Z-1727-2025 · Posted 2025-05-09

Recalling FirmMedical Depot Inc.
Recall StatusOpen, Classified
Product CodeCAW
Date Initiated2025-04-11
Product Quantity89,532 units

Reason for Recall

Potential for DC Power Supply housing to become hot to the touch and deform the plastic housing.

Firm Action

An URGENT NOTICE MEDICAL DEVICE RECALL dated 4/11/25 was emailed to consignees. The recall notification instructs consignees on how to identify affected devices in their possession, destroy the affected charger cord, and provides instructions on how to obtain a replacement charger cord through Drive Devilbiss. If product was further distributed, consignees are to contact their customers; Devilbiss has a draft recall notification and recall website available for their consignees who further distributed product. For those units that are rentals, Devilbiss recommends that consignees check devices as they return from rental use. Dan Hinkbein is available to assist consignees who distributed affected devices fulfill recall actions and can be contacted at [email protected]. Consignees with any issues or questions can call Drive Devilbiss Healthcare at 888-770-7017 Monday through Friday from 7:30 AM to 5:00 PM CST. Questions can also be emailed to [email protected]. The firm has set up a website for this recall event at www.recallrtr.com/iGo2-car-cord.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Albania, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, ISRAEL, Italy, JAPAN, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Maldives, Nepal, Netherlands, Norway, Oman, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovak Republic, Slovenia, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad & Tobago, Turkey, Uganda, Ukraine, United Arab Emirates.