Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Hc2 System Software Suite 4.0 version 3.0, available as component in Qiagen user guide, catalog #5050-00162, and as a component within a Qiagen PC, available as catalog # 5050-11140A. The software calculates and interprets assay results generated by the Digene Microplate Luminometer according to assay validation parameters. The software is used in conjunction with the following QIAGEN products: digene HC2 High-Risk HPV DNA Test [5199-1220 (one-plate); 5199-00016 (four-plate)] digene HC2 HPV DNA Test (5198-1220) digene HC2 CT/GC DNA Test (5130-1220) digene HC2 CT-ID DNA Test (5135-1220) digene HC2 GC-ID DNA Test (5140-1220) digene HC2 CT-GC Dual-ID DNA Test (5145-1220)

FDA Recall Z-1728-2011 · Posted None

Recall StatusTerminated
Product CodeJJQ
Date Initiated2011-01-12
Date Terminated2011-11-14
Product Quantity163

Reason for Recall

The software component of a diagnostic medical device may be defective and could cause incorrect patient values.

Firm Action

The firm, Qiagen, sent a "Notice of Product Field Action" letter dated January 12, 2011 to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue the process of running multiple assay protocols on a single microplate to eliminate the potential for inaccurate patient result reporting, and to complete and return the acknowledgment section via fax to QIAGEN 661-702-3854 or scan the document and email to [email protected]. The firm informed users that they will provide further detailed instructions for a workaround and software fix. A QIAGEN representative will contact the customers who indicated that they run multiple assays on a plate and are currently using the HC2 System Software Version 3.0 (Suite 4.0), to assist in the identification of those patients that may have been affected by the software defect. If you have any questions or concerns, please contact Qiagen Technical Services at 800-344-3631, Option 2.

Root Cause

Software design

Distribution Pattern

The recalled product was sold to medical laboratories nationwide and in Canada.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑