Recall StatusOpen, Classified
Date Initiated2025-04-08
Product Quantity45
Reason for Recall
Thrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.
Firm Action
On 4/8/2025, correction and removal notices were distributed to customers who were asked to do the following:
1) Review notice and the accompanying copy of the device IFU (IU-01087 Rev. C) and disseminate it to any applicable personnel and device users at your facility and network to ensure they are aware of the unused device replacement.
2) Notice should be shared with any organization where affected devices may have been transferred.
3) Complete and return the acknowledgement reply form via email to
If you have any questions, please contact your local sales representative, or the firm's Customer Service at 877-923-4747, or Quality department at
[email protected].
Root Cause
Labeling Change Control
Distribution Pattern
US Nationwide distribution in the states of FL, IL, WI, KY, TX, CA, OH, GA, AK, MI, PA, WA, CO, OH.