eTRAX Needle Sensor - 12G (for Aurora Trackers), Part Number 667-156
FDA Recall Z-1735-2026 · Posted 2026-04-01
Advertisement
Recalling FirmCivco Medical Instruments Co. Inc.
Recall StatusOpen, Classified
Product CodeIYO
Date Initiated2026-03-02
Product Quantity5 units
Reason for Recall
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.
Firm Action
An URGENT: MEDICAL DEVICE RECALL notification letter dated 3/2/26 was sent to customers.
Actions to be taken by the Customer/User:
CIVCO requests that you check your inventory for any eTRAX Needle Sensor for Aurora Trackers and complete the supplied response letter for the appropriate action.
Please contact your distribution partner to return any existing inventory, or report its destruction by completing the attached response form.
Please complete the enclosed response form even if you do not have any eTRAX Needle Sensor for Aurora Trackers remaining in your inventory to assist in our reconciliation process.
Response forms can also be sent via email to [email protected].
For questions, contact CIVCO at # 800-445-6741, or 319-248-6757.
Type of Action by the Company:
This notice is being communicated to all customers who have purchased an eTRAX Needle Sensor for Aurora Trackers from affected lots. CIVCO requests you provide this notification to the appropriate personnel within your facility. If the affected product was distributed outside of your organization, please notify those locations down to the medical facility level.
Root Cause
Software change control
Distribution Pattern
US States: OH, PA, WA. China.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.