Terminated

Verify SixCess Challenge Pack. Used to confirm that critical parameters of steam sterilization have been reached within a challenging test pack design.

FDA Recall Z-1736-2022 · Posted None

Recall StatusTerminated
Product CodeJOJ
Date Initiated2022-08-10
Date Terminated2024-10-30
Product Quantity619 boxes

Reason for Recall

Product contains incorrect chemical indicators in packaging.

Firm Action

The recalling firm issued customer and distributor letters beginning 08/10/2022 to impacted consignees. Consignees are instructed to: 1) Immediately destroy on-hand inventory of LCC003 Challenge Packs, Lot 3411228A 2) Complete the Medical Device Recall Response Form included with the notification letter and return the complete form via email to: [email protected] or via fax to 440-392-8963 3) STERIS will coordinate shipment of replacement product upon receipt of the completed Recall Response Form.

Root Cause

Process control

Distribution Pattern

Distribution in the US only - AK, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KY, LA, MA, MN, MO, MS, NE, NH, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, and WY