Unilateral external fixation ankle clamp Product Usage: Unilateral external fixation for use treatment of bone conditions amenable to treatment by use of external fixation modality
FDA Recall Z-1741-2015 · Posted 2015-06-09
Advertisement
Recalling FirmBiomet, Inc.
Recall StatusTerminated
Product CodeKTT
Date Initiated2015-05-01
Date Terminated2015-12-08
Product Quantity19
Reason for Recall
Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be necessary to adequately reconnect the fixator to the bone.
Firm Action
Biomet sent an URGENT MEDICAL DEVICE RECALL NOTICE letter dated May 1, 2015 to the affected distributors and implanting surgeons. The recall letter included a description of the reason for the recall, affected product, consignee responsibilities, and distributor instructions for responding to the formal recall notification. Consignees were instructed to return the affected product and continued monitoring of patients. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Root Cause
Process control
Distribution Pattern
US Nationwide Distribution in the states of CA, TX, FL, NC, MI, GA.