ARTERIAL LINE KIT, Custom Arterial Kit REF/REORDER NO. 498439 DISPOSABLE SINGLE, Rx. Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751.

FDA Recall Z-1744-2010 · Posted 2010-06-02

Recall StatusTerminated
Product CodeKGC
Date Initiated2010-03-29
Date Terminated2010-07-29
Product Quantity10 kits

Reason for Recall

There are 1-2 places on some of the multi-purpose kit blisters/trays in which there are holes and/or the material is so thin that holes are easily created. This would potentially cause the sterility of the products to be compromised.

Firm Action

Argon Medical Devices issued an "Urgent Medical Device Recall" notification via certified mail dated March 29, 2010 to distributors and end users. A Response Form was attached. Consignees were informed of the affected product and asked to quarantine and return all unused product to the firm. For further information, contact Argon Medical Devices at 1-903-677-9375.

Root Cause

Package design/selection

Distribution Pattern

Distributed to United States (IL, AL, NY, TX, OK, MI, NC, CA, OH, VA, KY, CO, LA, WA, MO, AND IA) and Canada.