Multi Vector Distractor Pin Holding Clamp, limited bone stock The holding clamp is part of the Multi Vector Distractor which cis used for mandibular bone lengthening, including conditions such as congenital mandibular deficiencies or post-traumatic defects. The Multi Vector Distractor can be utilized to perform bone transport procedures as an alternative to bone grafts and free flaps.
FDA Recall Z-1745-2013 · Posted 2013-07-17
Advertisement
Recalling FirmSynthes USA HQ, Inc.
Recall StatusTerminated
Product CodeDZL
Date Initiated2013-05-31
Date Terminated2015-08-31
Product Quantity121
Reason for Recall
Synthes is initiating a voluntary Medical Device removal for specified part numbers and lot numbers of the Mutis Vector Distractor Pin Holding clamp, limited bone stock. The firm received complaints that the clamp and nut construct would not engage the distractor arm threads and subsequently bind. It is not likely that a serious adverse event will result due to a nonconforming clamp, however there
Firm Action
Synthes sent a Urgent Notice: Medical Device Removal letter dated May 31, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers.
The firm received complaints that the clamp and nut construct would not engage the distractor arm threads and subsequently bind. It is not likely that a serious adverse event will result due to a nonconforming clamp, however, there is a remote potential that additional intervention would be required due to a non-functioning clamp.
Synthes requested the product to be removed from inventory and returned to Synthes by calling 1-800-479-6329 to obtain a Return Authorization Number. Direct Accounts were asked to complete and return the Verification Section at the end of the letter indicating the quantity of affected product was found. Negative responses were also requested.
If you have any questions please call (610) 719-5450.
Root Cause
Device Design
Distribution Pattern
US Nationwide Distribution including the states of MA, TX, MN, SC, UT, VA, and NY.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement