Recall StatusOpen, Classified
Date Initiated2021-11-21
Product Quantity93 US; 57 OUS
Reason for Recall
During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.
Firm Action
Customers were notified of the initial issue by email beginning Nov 3 2021, and
requested to return acknowledgment of the notice.
On January 26, 2022, a new version of the protocol software was released and customers were advised of the new protocol availability and acknowledge download via a follow up notice.
Distribution Pattern
Domestic distribution to AZ, CA, CO, DE, DC,
FL, GA, IA, KS, LA,
MA, MD, MN, NV, NJ,
NY, NC, OH, PA, SC,
SD, TN, TX, VA, WA,
WI.
International distribution worldwide.