Boston Scientific, NovaGold High Performance Guidewire, Rx only, Sterilized using ethylene oxide, UPN M00552010 and M00552000, Manufactured by NeoMetrics Inc, 2605 Fernbrook Lane Suite J, Plymouth, MN 55447. The NovaGold Guidewire is intended for use in selective cannulation of the biliary ducts including the common bile, pancreatic, cystic, right and left hepatic ducts, and to aid in the placement of diagnostic and therapeutic devices during endoscopic procedures.

FDA Recall Z-1747-2014 · Posted 2014-06-10

Recalling FirmNEO METRICS, INC.
Recall StatusTerminated
Product CodeOCY
Date Initiated2014-03-31
Date Terminated2015-01-16
Product Quantity1256 boxes ( 2 wires/box)

Reason for Recall

Neo Metrics, Inc. is recalling NovaGold Guidewire because the tip may uncoil and / detach more frequently than anticipated. There have been no reports of illness or injury.

Firm Action

NeoMetrics sent a letter dated March 31, 2014, to Boston Scientific. On March 24, 2014, Boston Scientific made NeoMetrics aware that Boston Scientific received 3 complaints from customers that the tip of the NovaGold guidewire detached and/or became uncoiled. Because the tips are breaking and/or uncoiling at a rate that was not anticipated, it has been decided that these units should be removed from the field. Therefore, NeoMetrics does not object to Boston Scientific executing this product removal in accordance with Section 4.4 of the distribution agreement. For further questions please call (763) 559-4440.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - US and the countries of Austria, Belgium, Germany, Spain, France, Great Britain, Italy, Netherlands and Sweden.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑