FoundationOne Companion Diagnostic (F1CDx)
FDA Recall Z-1747-2025 · Posted 2025-05-12
Advertisement
Recalling FirmFoundation Medicine, Inc.
Recall StatusOpen, Classified
Product CodePQP
Date Initiated2025-04-09
Product Quantity4 units
Reason for Recall
Four (4) total reports were sent to customers with clinically relevant variants requiring amendment, one (1) of which with an incorrect negative CDx claim. The remaining three (3) with tumor profiling result changes did not impact CDx indications.
Firm Action
On April 9, 2025 URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters were sent to customers.
Asked Customer Immediate Actions:
Foundation Medicine respectfully requests your support with the following actions:
1. Review, complete, sign and return the attached business reply form to be returned in
accordance with the directions on the business reply form within 5 business days of
receipt of this notification.
2. Forward this notice to anyone in your facility that needs to be informed.
3. If there is any impact that requires this notice to be forwarded to another facility, contact
that facility and provide them with this letter.
4. Keep a copy of this notice.
Adverse reactions or quality problems experienced with the use of this product may be reported
to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by
fax.
Foundation Medicine apologizes for any inconvenience caused by this issue. If any further
information or support concerning this issue is needed, please contact Foundation Medicine at
any time.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide - US Nationwide distribution in the states of GA, NY and the country of Russia.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.