Terminated

PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 90110, 100 test kit (obsolete), manufactured by DiaDexus, South San Francisco, CA.

FDA Recall Z-1749-2010 · Posted 2010-06-03

Recalling FirmDiaDexus, Inc
Recall StatusTerminated
Product CodeNOE
Date Initiated2009-10-29
Date Terminated2011-01-26
Product Quantityall lots

Reason for Recall

Inaccurate results-- some samples may return with falsely elevated results potentially as a result of an interfering substance. Some patients may be subjected to additional lifestyle changes or intensified pharmaceutical therapy; such as statins, as a result of the falsely elevated result.

Firm Action

Notification letter issued to consignees on 10/29/2009. Customers were instructed to not report PLAC test results for samples that are greater than 360 ng/mL, and pursue confirmatory testing by contacting DiaDexus. Customers with questions please call the company technical support line toll free at 1-877-752-2837 for assistance. UPDATE: On April 30, the firm suspended marketing of the PLAC TIA method, as the limited range no longer covered the clinical data/reference range data cited in the product insert. The firm sent out a letter notifying customers on May 10, 2010. The firm set up a sample send out program, so that customers can send samples to labs running the ELISA method PLAC test. Arrangements were made so that labs could be enabled to run a CLIA complex assay, for customer credit programs for un-used products, and a fax back response form for response tracking.

Root Cause

Other

Distribution Pattern

Product was distributed to 123 consignees throughout the US.