Terminated

Omnifit EON Surgical Protocol; Literature Number: LSP48 4/04 Stryker Howmedica Osteonics Corp., Mahwah, NJ

FDA Recall Z-1750-2008 · Posted 2008-08-17

Recall StatusTerminated
Product CodeLXH
Date Initiated2007-05-17
Date Terminated2008-08-19
Product Quantity2501 units

Reason for Recall

The Table 2 Broach and Neck Trial Sizing (Literature # LSP48 04/04) on page 3 is incorrect, reading 5mm less than what it should be for certain sizes.

Firm Action

Styrker Branches/Agencies were notified of the incorrect information printed in the protocal on May 17, 2007 via an Important Product Correction letter. The letter requested that all incorrect copies of the protocol be located and destroyed immediately. They requested that the Acknowledgement Form be completed and faxed, and that a Stryker Orthopaedics Customer Service rep be contacted in order to reorder the destroyed product.

Root Cause

Labeling design

Distribution Pattern

Worldwide Distribution --- including USA and countries of Canada, Australia, Latin America and Greece.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑