Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system

FDA Recall Z-1750-2013 · Posted 2013-07-17

Recall StatusTerminated
Product CodeIZI
Date Initiated2013-01-31
Date Terminated2014-11-12
Product Quantity23

Reason for Recall

During the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail.

Firm Action

Siemens sent an Urgent Customer Information letter to their consignees. The letter identified the affected product, problem , update Instructions and actions to be taken. Siemens began to equip affected systems with a modified board assembly in January, 2013.

Root Cause

Device Design

Distribution Pattern

United States Nationwide Distribution - including the states of: CA, CO, FL, IA, IL, IN, MA, MD, MI, NC, NY, PA, TX, VA and WA and Puerto Rico.