Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body
FDA Recall Z-1750-2018 · Posted None
Advertisement
Recalling FirmMedtronic Sofamor Danek USA Inc
Recall StatusTerminated
Product CodeKWQ
Date Initiated2018-03-23
Date Terminated2021-07-25
Product Quantity39 units
Reason for Recall
Slot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm.
Firm Action
The firm initiated their recall by letter on 03/23/2018. The recall notice requested:
"1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return via email to [email protected].
3.) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product. "
Root Cause
Process control
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of GA, CT, LA, MD, WY, IN, MD, FL, and France
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.