Hydra Vision DR 60/80 System, Models 700539 and 700540, Liebel-Flarshiem Company. The System is used in radiologic imaging procedures.
FDA Recall Z-1753-2013 · Posted 2013-07-18
Advertisement
Recalling FirmLeibel-Flarsheim Company LLC
Recall StatusTerminated
Product CodeOWB
Date Initiated2013-04-25
Date Terminated2013-07-26
Product Quantity12 units
Reason for Recall
A complaint was received in which a hospital physicist indicated the dosage was too high on the HUT system. During the service call the systems Maximum Fluoro Dose was found to be 30 R/min which is three times the Maximum Fluoro Dose limit of 10 R/min.
Firm Action
The customers received an "Urgent Device Correction" notification letter on May 12, 2013 explaining the product issue and providing recommended actions. Customers were contacted to set up a service call to test and recalibrate equipment as necessary. Questions were directed to 800-778-7898.
Root Cause
Employee error
Distribution Pattern
Nationwide Distribution (US), including the states of TN, KY, CO, KS, CT, HI, and AZ.