INNOMED CUP REMOVAL STARTER INSTRUMENT, REF 520054, Hip cutter blade cup
FDA Recall Z-1753-2024 · Posted 2024-05-01
Advertisement
Recalling FirmInnomed, Inc.
Recall StatusCompleted
Product CodeHWB
Date Initiated2024-03-07
Product Quantity41 devices
Reason for Recall
54 MM finisher blade was inadvertently affixed to the shaft of the CupX handle in lieu of the 54 MM starter blade.
Firm Action
Innomed issued an urgent medical device recall notice to its consignees on 03/07/2024 via email. The notice explained the issue, potential risk to health, and requested the return of the affected devices for US distribution and destruction of the affected devices distributed outside the US.
Root Cause
Under Investigation by firm
Distribution Pattern
FL, GA, PA, NY, CA, VA, SD, MN, TX, UT, NC, and UK, Denmark, Germany Switzerland, Ireland, Canada