The Baylis LumbarCool Pain Management Kit consisting of: 1 LumbarCool Pain Management Probe, 3 LumbarCool Pain Management introducers, and 1 Pain Management Tube Kit. Model/Catalogue Number: LUK-17-150-4 The LumbarCool Pain Management System, a cooled radiofrequency pain management system, in combination with the Baylis Pain Management Generator-TD (PMG-TD), is indicated for use to create RF lesions in nervous tissue.
FDA Recall Z-1755-2010 · Posted 2010-06-03
Advertisement
Recalling FirmBaylis Medical Corp *
Recall StatusTerminated
Product CodeGXI
Date Initiated2010-05-12
Date Terminated2010-10-04
Product Quantity58
Reason for Recall
Name of the device reflected on the product packing sleeve is incorrect.
Firm Action
The firm, Baylis Medical, notified US Distributor by email and sent a "URGENT Field Corrective Action Notice" dated May 12, 2010, to customer. The notifications described the product, problem and action to be taken by customer. The customers were given the follow options: Option 1: In order to prevent confusion in identification of the product before use, please immediately remove and dispose the package sleeve that covers the tray packaging of the LumbarCool Pain Management Kit.
Option 2: Alternatively, you may exchange your kit from Lot LKFA160310 with a correctly labeled replacement LumbarCool Pain Management Kit by contacting Kimberly-Clark Health Care sales representative. The customers were ask to fill out the Field Corrective Action Acknowledgment and return it.
Any question please call Baylis Sales Representative at 905-602-4875 and/or contact your local sales representative for any further information you may require.
Root Cause
Error in labeling
Distribution Pattern
Nationwide distribution: TX
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.