AEQUALIS HUMERAL NAIL DRILL BIT, REF 7020140 Product Usage: The Aequalis Humeral Nail system is intended to provide temporary stabilization of various types of proximal and/or diaphyseal fractures of the humerus. Types of fractures include, but are not limited to, non-unions, malunions, malalignments, pathological fractures, and impending pathological fractures. Examples of specific indications according to AO classification include Type A-Fractures, dislocated, Type B Fractures, dislocated, Type C-Fractures, with intact humeral head, or Humeral Fractures according to Neer-Classification (2, 3 and 4 part fractures).
FDA Recall Z-1757-2018 · Posted None
Advertisement
Recalling FirmTornier, Inc
Recall StatusTerminated
Product CodeHTW
Date Initiated2018-03-26
Date Terminated2021-08-12
Product Quantity3327 units
Reason for Recall
Drill bit tips without new BargerGard tip protector may puncture through sterile packaging.
Firm Action
The firm contacted its distributors by email on 26-March-2018, and a written letter sent 27-March-2018 via FedEx. The Medical Facility was contacted with a written letter sent 28-March-2018 via FedEx.. The letter identified the affected product, problem and actions to be taken. Customers were asked to complete and return the attached Acknowledgement to Wright Medical via email to [email protected]. For questions call 952-683-7482.
Root Cause
Package design/selection
Distribution Pattern
US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.