AEQUALIS HUMERAL NAIL DRILL BIT, REF 7020140 Product Usage: The Aequalis Humeral Nail system is intended to provide temporary stabilization of various types of proximal and/or diaphyseal fractures of the humerus. Types of fractures include, but are not limited to, non-unions, malunions, malalignments, pathological fractures, and impending pathological fractures. Examples of specific indications according to AO classification include Type A-Fractures, dislocated, Type B Fractures, dislocated, Type C-Fractures, with intact humeral head, or Humeral Fractures according to Neer-Classification (2, 3 and 4 part fractures).

FDA Recall Z-1757-2018 · Posted None

Recalling FirmTornier, Inc
Recall StatusTerminated
Product CodeHTW
Date Initiated2018-03-26
Date Terminated2021-08-12
Product Quantity3327 units

Reason for Recall

Drill bit tips without new BargerGard tip protector may puncture through sterile packaging.

Firm Action

The firm contacted its distributors by email on 26-March-2018, and a written letter sent 27-March-2018 via FedEx. The Medical Facility was contacted with a written letter sent 28-March-2018 via FedEx.. The letter identified the affected product, problem and actions to be taken. Customers were asked to complete and return the attached Acknowledgement to Wright Medical via email to [email protected]. For questions call 952-683-7482.

Root Cause

Package design/selection

Distribution Pattern

US Nationwide Distribution