Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.

FDA Recall Z-1759-2024 · Posted 2024-05-03

Recall StatusOpen, Classified
Product CodeJIX
Date Initiated2024-03-05
Product Quantity80,366 units (US: 242 units; OUS: 80,094 units)

Reason for Recall

Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.

Firm Action

Consignees were emailed a Medical Device Correction notice dated 3/5/2024 instructing them to discuss the contents of the notification with their facility's Medical Director if affected units have been used, replace the IFU for affected devices with the one provided with the notification, and complete and return the provided response form by email to technical [email protected] within five working days. Updated IFUs can be accessed at www.randox.com and questions can be directed to [email protected].

Root Cause

Process design

Distribution Pattern

Domestic: Nationwide Distribution.