Hudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5-10213 3) CF 7.0 mm, Product Codes: 5-10114, 5-10214 4) CF 7.5 mm, Product Codes: V5-10115, 5-10115, 5-10215 5) CF 8.0 mm, Product Codes: 5-10116, 5-10216, V5-10116 6) CF 8.5 mm, Product Codes: 5-10117, 5-10217 Product Usage: Tracheal tube/airway management
FDA Recall Z-1760-2019 · Posted 2019-06-19
Advertisement
Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeBTR
Date Initiated2019-05-24
Date Terminated2022-02-15
Product Quantity1101073 units
Reason for Recall
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
Firm Action
Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product.
Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex..
For questions contact your local sales representative or Customer Service at 1-866-396-2111
Root Cause
Process control
Distribution Pattern
Worldwide - US Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.