Terminated

Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-22515 5) EZ-ENDO 8.0 mm, Product Code 5-22516 6) EZ-ENDO 8.5 mm, Product Code 5-22517 Product Usage: Tracheal tube/airway management

FDA Recall Z-1761-2019 · Posted 2019-06-19

Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeBTR
Date Initiated2019-05-24
Date Terminated2022-02-15
Product Quantity55312 units

Reason for Recall

This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.

Firm Action

Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111

Root Cause

Process control

Distribution Pattern

Worldwide - US Nationwide Distribution