DURAMAX (TM) Hemodialysis Catheters, Stacked Tip 24 cm Straight Basic Kit, Catalog Number/REF 10302803, 510 (k) 080400, Sterile --- Manufactured in USA by AngioDynamics, Inc. 603 Queensbury Avenue, Queensbury, NY 12804

FDA Recall Z-1762-2010 · Posted 2010-06-04

Recalling FirmAngiodynamics, Inc.
Recall StatusTerminated
Product CodeMSD
Date Initiated2010-01-08
Date Terminated2012-07-25
Product Quantity675

Reason for Recall

The product has the potential for a cross-lumen leak. One of the two leak testers that are used to perform the leak test on the DuraMax Catheters was found to be faulty. The tester was not able to detect cross-lumen leaks. Both testers were used during the manufacturing 100% leak test.

Firm Action

Customers were sent URGENT MEDICAL DEVICE RECALL LETTERS (dated 1/07/10) via Certified Mail, Return Receipt Requested. Customers were instructed to return all recalled product. A return authorization number was assigned to each customer in the recall notification letter and a UPS account was provided for return freight. A reply form was provided in the recall notification letter. Questions were directed to the company sales representatives or to Customer Service at 800-772-6446.

Root Cause

Other

Distribution Pattern

Nationwide and St. Thomas, VI