NanoSight LM10 fitted with LM14 module. A laboratory instrument for determination of particle characteristics (e.g. size).
FDA Recall Z-1762-2022 · Posted 2022-09-26
Advertisement
Recalling FirmMALVERN PANALYTICAL LTD
Recall StatusOpen, Classified
Product CodeREN
Date Initiated2022-02-28
Product Quantity76 units
Reason for Recall
Product found to not comply to Class 1 Laser safety requirement.
Firm Action
Customers are being made aware of this issue. Malvern Panalytical will retrofit modified microscope plates and cell tops with a new safety cover on the LM10 to maintain the original Class 1 laser product classification at no charge to the purchaser.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.